91* For a smaller intended dose of 0.05 mL, the same 0.070 mL dead space exceeds the dose entirely; no plunger position can deliver 0.05 mL if that volume of air must first be displaced from the hub.91* For a smaller intended dose of 0.05 mL, the same 0.070 mL dead space exceeds the dose entirely; no plunger position can deliver 0.05 mL if that volume of air must first be displaced from the hub.
9292
+93The vial-yield arithmetic follows from the wastage figure. Each dose consumes 0.20 mL delivered plus 0.070 mL retained, or 0.270 mL. From 2.00 mL, seven full doses can be drawn (7 × 0.270 = 1.89 mL), with 0.11 mL remaining — insufficient for an eighth. The nominal ten doses have become seven, a 30% reduction. With a fixed-needle insulin syringe at 0.005 mL, each dose consumes 0.205 mL and nine full doses can be drawn, with 0.155 mL remaining.{{r|ppdeadspace}}
+94
93=== Priming and air management ===95=== Priming and air management ===
94The mechanism by which the dose error arises is the presence of gas where liquid is assumed. Descriptions of injection technique in manufacturers' [[Prescribing information|prescribing information]] and in nursing references accordingly treat expulsion of air before the dose is measured as a defined step, and treat the graduation as valid only for a gas-free barrel and hub.{{r|who2015}}96The mechanism by which the dose error arises is the presence of gas where liquid is assumed. Descriptions of injection technique in manufacturers' [[Prescribing information|prescribing information]] and in nursing references accordingly treat expulsion of air before the dose is measured as a defined step, and treat the graduation as valid only for a gas-free barrel and hub.{{r|who2015}}