1{{Infobox concept1{{Infobox concept
2| name = Syringe dead space2| name = Syringe dead space
+3| image = syringe.svg
+4| caption = Dead space lies distal to the plunger tip at full depression: the hub cavity, the needle lumen and any recess in the plunger face.
3| Also known as = Residual volume; hold-up volume; wastage volume5| Also known as = Residual volume; hold-up volume; wastage volume
4| Location = Needle hub, needle lumen, plunger-tip recess6| Location = Needle hub, needle lumen, plunger-tip recess
60| 30G | 0.312 mm | 0.159 mm | 0.25 µL |62| 30G | 0.312 mm | 0.159 mm | 0.25 µL |
61| 31G | 0.261 mm | 0.133 mm | 0.18 µL |63| 31G | 0.261 mm | 0.133 mm | 0.18 µL |
+64
+65Volumes are computed as the product of cross-sectional area and length; outside diameters follow the metric equivalents of the Birmingham gauge series used in ISO 9626, and inside diameters are typical regular-wall values.{{r|iso9626}} A 27G half-inch needle therefore holds 0.44 µL, roughly one part in 160 of a 70 µL hub. Changing needle gauge or length has almost no effect on dead space; changing the hub design changes it by an order of magnitude.