PeptidePedia The community reference

Retention time (revision 9)

Old revision·22:28, 21 Dec 2024·CiteBot

This is an old revision of this page, as it stood at 22:28, 21 Dec 2024, saved by CiteBot with the summary bot: repair redlinked category. It may differ substantially from the current revision, and any error it contains may since have been corrected.
Retention timeChromatography
048121620main peak 12.4 min · 98.7 area%mAUminutes
Symbolt_R
Depends onColumn, mobile phase, gradient, flow, temperature
Relative measureRelative retention time, against a reference peak
Topic infobox · conventions

Retention time is the interval between injection and the appearance of a component's peak maximum at the detector. Under fixed conditions it is characteristic of a substance and reproducible, which makes it the basis of chromatographic identification.[1]

It is not a property of the substance alone. Column chemistry and dimensions, mobile-phase composition, gradient profile, flow rate, temperature and even instrument dwell volume all change it, so a retention time is only meaningful alongside the conditions that produced it.[1]

Relative retention time — the ratio of a peak's retention to that of a reference peak in the same run — is more robust, because system variations that shift both peaks cancel.[2][3]

What it can identify

[edit]

Matching a sample peak's retention to that of a reference standard run under identical conditions on the same day is the ordinary chromatographic identification. It is weak evidence on its own: many species can share a retention time on a given system.[2]

Co-injection strengthens it substantially. Spiking the sample with the standard should enlarge the existing peak without producing a shoulder; a shoulder indicates two species.

The strongest form combines retention matching with an orthogonal determination — mass confirmation or a second chromatographic mode with different selectivity. This is what pharmacopoeial identification tests specify.[1]

Drift and its causes

[edit]
CauseEffect
Column ageingProgressive shift, usually earlier
Temperature changeShift in either direction
Mobile-phase preparationSmall shifts, run to run
Instrument dwell volumeSystematic offset between instruments
Gradient delaySame, for gradient methods

Dwell-volume differences are the reason a method transferred between instruments often needs adjustment: the delay between the pump forming a gradient and it reaching the column differs, so retention shifts even with identical settings.[1]

References

  1. ^ a b c d United States Pharmacopeia, General Chapter <621>, Chromatography.
  2. ^ a b United States Pharmacopeia, General Chapter <1503>, Quality Attributes of Synthetic Peptide Drug Substances.
  3. ^ Mant CT, Hodges RS. "Analysis of peptides by high-performance liquid chromatography." Methods in Enzymology 271:3–50 (1996). PMID 8782429.