PeptidePedia The community reference

Retention time (revision 10)

Old revision·18:51, 6 Jan 2025·SatietySunniva

This is an old revision of this page, as it stood at 18:51, 6 Jan 2025, saved by SatietySunniva with the summary expand §What it can identify. It may differ substantially from the current revision, and any error it contains may since have been corrected.
Retention timeChromatography
048121620main peak 12.4 min · 98.7 area%mAUminutes
Symbolt_R
Depends onColumn, mobile phase, gradient, flow, temperature
Relative measureRelative retention time, against a reference peak
Topic infobox · conventions

Retention time is the interval between injection and the appearance of a component's peak maximum at the detector. Under fixed conditions it is characteristic of a substance and reproducible, which makes it the basis of chromatographic identification.[1]

It is not a property of the substance alone. Column chemistry and dimensions, mobile-phase composition, gradient profile, flow rate, temperature and even instrument dwell volume all change it, so a retention time is only meaningful alongside the conditions that produced it.[1]

Relative retention time — the ratio of a peak's retention to that of a reference peak in the same run — is more robust, because system variations that shift both peaks cancel.[2][3]

What it can identify

[edit]

Matching a sample peak's retention to that of a reference standard run under identical conditions on the same day is the ordinary chromatographic identification. It is weak evidence on its own: many species can share a retention time on a given system.[2]

Co-injection strengthens it substantially. Spiking the sample with the standard should enlarge the existing peak without producing a shoulder; a shoulder indicates two species.

The strongest form combines retention matching with an orthogonal determination — mass confirmation or a second chromatographic mode with different selectivity. This is what pharmacopoeial identification tests specify.[1]

Drift and its causes

[edit]
CauseEffect
Column ageingProgressive shift, usually earlier
Temperature changeShift in either direction
Mobile-phase preparationSmall shifts, run to run
Instrument dwell volumeSystematic offset between instruments
Gradient delaySame, for gradient methods

Dwell-volume differences are the reason a method transferred between instruments often needs adjustment: the delay between the pump forming a gradient and it reaching the column differs, so retention shifts even with identical settings.[1]

Because of all this, a retention time quoted on a certificate is a within-system value. It is worth reporting — it lets a repeat determination be checked for gross discrepancy — but it is not a transferable identifier.[2]

See also

References

  1. ^ a b c d United States Pharmacopeia, General Chapter <621>, Chromatography.
  2. ^ a b c United States Pharmacopeia, General Chapter <1503>, Quality Attributes of Synthetic Peptide Drug Substances.
  3. ^ Mant CT, Hodges RS. "Analysis of peptides by high-performance liquid chromatography." Methods in Enzymology 271:3–50 (1996). PMID 8782429.