| + | 30 | == Reading an LC-MS result == | |
| + | 31 | The fields that make an LC-MS determination interpretable are the chromatographic conditions, the ionisation mode, the mass analyser and its resolution, and whether the reported mass is monoisotopic or average.{{r|aebersold2003}} | |
| + | 32 | ||
| + | 33 | Resolution determines what could have been distinguished: a 0.984 Da [[Deamidation|deamidation]] shift is invisible on a unit-resolution instrument, while a 16 Da oxidation is not. A report omitting the instrument omits the answer to what it could have seen.{{r|usp1503}} | |
| + | 34 | ||