Electrospray ionisation transfers molecules from solution into the gas phase as ions. A solution is sprayed from a capillary held at high potential, producing charged droplets whose solvent evaporates until ions are released.[1]

It is a soft technique — the molecule survives largely intact — and it produces multiply charged ions from peptides and proteins, so a large molecule appears at a mass-to-charge ratio within the range of ordinary analysers. A single peptide therefore produces a series of peaks at successive charge states, deconvolved by software into one neutral mass.[2]

Because it works from flowing solution, it couples directly to chromatography, which is what makes LC-MS possible.[1]

Mechanism

[edit]

The applied potential draws the liquid into a cone from whose tip a fine spray of charged droplets emerges. Solvent evaporates, the droplets shrink until charge repulsion exceeds surface tension, and they divide. Ions are ultimately released either by field-driven ejection from very small droplets or by complete evaporation of the solvent around them.[1]

The charge states observed depend on the number of ionisable sites and on solution conditions. Acidic mobile phases favour protonation and positive-mode operation, which is why peptide LC-MS is usually run in positive mode with an acidic modifier.[2]

Ion suppression is the characteristic difficulty: co-eluting species compete for charge, so a compound's signal depends on what else is eluting at the same time. This is why the technique is not quantitative without calibration.[3]