| 87 | A worked case. A vial containing 5.0 mg of peptide is reconstituted with 2.00 mL of diluent, giving 2.50 mg·mL⁻¹. An intended dose of 0.50 mg corresponds to 0.20 mL. | 87 | A worked case. A vial containing 5.0 mg of peptide is reconstituted with 2.00 mL of diluent, giving 2.50 mg·mL⁻¹. An intended dose of 0.50 mg corresponds to 0.20 mL. |
| 88 | 88 | ||
| + | 89 | * With a device of 0.070 mL dead space and an unexpelled hub air pocket of that size, the liquid delivered is 0.20 − 0.07 = 0.13 mL, containing 0.325 mg. The shortfall is 35% of the intended dose. | |
| + | 90 | * With the same device correctly primed and free of air, 0.20 mL and 0.50 mg are delivered, and 0.070 mL — 0.175 mg — is discarded in the device. | |
| + | 91 | * For a smaller intended dose of 0.05 mL, the same 0.070 mL dead space exceeds the dose entirely; no plunger position can deliver 0.05 mL if that volume of air must first be displaced from the hub. | |
| + | 92 | ||