PeptidePedia The community reference

Sterility testing (revision 10)

Old revision·04:21, 7 Feb 2025·AnalyticalAnnie

This is an old revision of this page, as it stood at 04:21, 7 Feb 2025, saved by AnalyticalAnnie with the summary typo. It may differ substantially from the current revision, and any error it contains may since have been corrected.
Sterility testingCompendial test
Compendial chapterUSP General Chapter 71
MethodsMembrane filtration; direct inoculation
IncubationNot less than 14 days
Nature of resultStatistical, not absolute
Analytical method infobox · conventions

Sterility testing determines whether a sample yields microbial growth when incubated in defined media under defined conditions. A passing result does not establish that a batch is sterile; it establishes that the units tested did not yield growth, which is a statistical statement about the batch rather than a property of it.[1]

This limitation is intrinsic and well recognised. Testing a small number of units from a large batch has limited power to detect low-level contamination, and for this reason sterility assurance in manufacture rests on process validation and environmental control rather than on end-product testing.[2]

Sterility is distinct from freedom from pyrogens. A sterile preparation may contain bacterial endotoxin from organisms killed during processing, and the two determinations are not substitutes for one another.[3]

Method

[edit]

Two methods are recognised. Membrane filtration passes the sample through a 0.45 μm filter that retains organisms, washes away any inhibitory substance, and incubates the membrane in growth media; it is preferred where the sample can be filtered. Direct inoculation adds the sample to media directly and is used where filtration is impractical.[1]

Two media are used in parallel: a soybean–casein digest medium incubated at 20–25 °C for fungi and aerobic organisms, and a fluid thioglycollate medium at 30–35 °C for anaerobes and aerobes. Incubation is for not less than fourteen days, which is why a sterility result is never available quickly.

Method suitability must be demonstrated before the test means anything. The sample must be shown not to inhibit growth of specified challenge organisms in the media used; a sample with antimicrobial activity that has not been neutralised will pass the test by suppressing the growth it is meant to reveal.[1]

Statistical limits

[edit]

The number of units tested is small — commonly ten to twenty depending on batch size — and contamination in a batch is usually sporadic rather than uniform. The probability of detecting a low contamination rate is correspondingly low.[2]

Contamination rate in batchApproximate chance of detection, 20 units tested
0.1%Very low
1%Low
10%Moderate
50%High

See also

References

  1. ^ a b c United States Pharmacopeia, General Chapter <71>, Sterility Tests.
  2. ^ a b United States Pharmacopeia, General Chapter <1211>, Sterility Assurance.
  3. ^ United States Pharmacopeia, General Chapter <85>, Bacterial Endotoxins Test.