26Gradient duration is the parameter with the largest effect on a reported purity figure. A shallow gradient run over an hour resolves species a fifteen-minute gradient merges, and the merged determination returns a higher number on identical material. Neither is wrong; they are different measurements. See [[Purity claim inflation]].{{r|mant1996}}26Gradient duration is the parameter with the largest effect on a reported purity figure. A shallow gradient run over an hour resolves species a fifteen-minute gradient merges, and the merged determination returns a higher number on identical material. Neither is wrong; they are different measurements. See [[Purity claim inflation]].{{r|mant1996}}
2727
+28Temperature is the parameter most often omitted from certificates. Elevated column temperature sharpens peaks for large peptides by reducing conformational heterogeneity, and a method developed at 60 °C will not reproduce at ambient.