Retention time
Symbolt_R
Depends onColumn, mobile phase, gradient, flow, temperature
Relative measureRelative retention time, against a reference peak
Topic infobox · conventions

Retention time is the interval between injection and the appearance of a component's peak maximum at the detector. Under fixed conditions it is characteristic of a substance and reproducible, which makes it the basis of chromatographic identification.[1]

It is not a property of the substance alone. Column chemistry and dimensions, mobile-phase composition, gradient profile, flow rate, temperature and even instrument dwell volume all change it, so a retention time is only meaningful alongside the conditions that produced it.[1]

What it can identify

[edit]

Matching a sample peak's retention to that of a reference standard run under identical conditions on the same day is the ordinary chromatographic identification. It is weak evidence on its own: many species can share a retention time on a given system.[2]

Co-injection strengthens it substantially. Spiking the sample with the standard should enlarge the existing peak without producing a shoulder; a shoulder indicates two species.

References

  1. ^ a b United States Pharmacopeia, General Chapter <621>, Chromatography.
  2. ^ United States Pharmacopeia, General Chapter <1503>, Quality Attributes of Synthetic Peptide Drug Substances.