Area percent purity (revision 2)
Old revision·05:10, 29 Aug 2024·KarlFischerKit
| Area percent purity | |
|---|---|
| Analytical method infobox · conventions |
Area percent purity is the integrated area of the principal peak in a chromatogram expressed as a percentage of the total integrated area of all peaks. It is the figure almost always meant when a research-peptide certificate reports "purity", and it is reported without further qualification often enough that its limitations are worth stating plainly.[1]
The measure has one defining property from which everything else follows: it is relative to what was detected, not to what was weighed. A species that does not elute, does not absorb at the detection wavelength, or co-elutes with the principal peak contributes nothing to the denominator and therefore cannot reduce the figure. Water, counterion and inorganic residue are invisible to it in the ordinary case, which is why a vial can honestly carry a certificate reading 99.2% and contain substantially less than 99.2% peptide by mass.[1]
How the figure is produced
[edit]A sample is dissolved, injected onto a reverse-phase column, and eluted under a gradient of increasing organic modifier. A detector — most often ultraviolet absorbance at 214 nm, where the peptide bond absorbs — produces a signal against time. Peaks are integrated, and the principal peak's area is divided by the sum of all integrated areas.
Four choices in that description change the answer, and none of them is usually stated on a certificate: